We have developed a biomagnetic measurement system for mouse by using high-Tc SQUID magnetometer. The system consists of a high-T c SQUID magnetometer, a table-top magnetically shielded box, and a nonmagnetic dewar with a warm-to-cool distance of about 4 mm. Mice with weights of about 27 g were anesthetized and fixed in supine position on a plastic xyz-stage inside the shield box. The SQUID readout signal was digitized in 24-bit and recorded with sampling rate of 1280 Hz. We carried out a scanning measurement by changing x-and y-position of the subject inside the shield box. Finally, we could obtain clear mouse MCG waveforms of about 20 pT in peak amplitude with well defined P-, QRS-, and T-wave. We demonstrated the high-Tc SQUID magnetometer based biomagnetic measurement system is useful for monitoring mouseMCG signals with sufficient noise-to-signal ratio.
Keyword
Biomagnetic measurementHigh-Tc SQUID magnetometerMagnetocardiographyMouse
ISSN
1051-8223
Publisher
IEEE
DOI
http://dx.doi.org/10.1109/TASC.2010.2094993
Type
Article
Appears in Collections:
Ochang Branch Institute > Division of National Bio-Infrastructure > Laboratory Animal Resource & Research Center > 1. Journal Articles
Files in This Item:
  • There are no files associated with this item.


Items in OpenAccess@KRIBB are protected by copyright, with all rights reserved, unless otherwise indicated.