Development of a rat biomagnetic measurement system using a high-TC SQUID magnetometer

Cited 3 time in scopus
Metadata Downloads
Title
Development of a rat biomagnetic measurement system using a high-TC SQUID magnetometer
Author(s)
I S Kim; Chul Ho Lee; Y H Lee
Bibliographic Citation
Superconductor Science & Technology, vol. 23, no. 8, pp. 085001-085001
Publication Year
2010
Abstract
We have developed a rat magnetocardiograph (MCG) system employing a high-TC SQUID magnetometer and a tabletop magnetic shield. We obtained clear MCG signals from a healthy Wistar Kyoto rat with a relatively high peak amplitude of 50pT by virtue of the small gap cryostat developed in this study. Well defined P-, QRS-and T-waves were observed on the MCG of the healthy rat. In the case of a spontaneously hypertensive rat measurement, the MCG showed quite a disturbed wave pattern thought to be caused by the hypertensive heart abnormality. The results suggest that the rat biomagnetic measurement system has a strong potential for monitoring the progress of the heart disease model.
ISSN
0953-2048
Publisher
IOP Publishing Ltd
DOI
http://dx.doi.org/10.1088/0953-2048/23/8/085001
Type
Article
Appears in Collections:
Ochang Branch Institute > Division of National Bio-Infrastructure > Laboratory Animal Resource & Research Center > 1. Journal Articles