- Title
- Development of a rat biomagnetic measurement system using a high-TC SQUID magnetometer
- Author(s)
- I S Kim; Chul Ho Lee; Y H Lee
- Bibliographic Citation
- Superconductor Science & Technology, vol. 23, no. 8, pp. 085001-085001
- Publication Year
- 2010
- Abstract
- We have developed a rat magnetocardiograph (MCG) system employing a high-TC SQUID magnetometer and a tabletop magnetic shield. We obtained clear MCG signals from a healthy Wistar Kyoto rat with a relatively high peak amplitude of 50pT by virtue of the small gap cryostat developed in this study. Well defined P-, QRS-and T-waves were observed on the MCG of the healthy rat. In the case of a spontaneously hypertensive rat measurement, the MCG showed quite a disturbed wave pattern thought to be caused by the hypertensive heart abnormality. The results suggest that the rat biomagnetic measurement system has a strong potential for monitoring the progress of the heart disease model.